You can not select more than 25 topics Topics must start with a letter or number, can include dashes ('-') and can be up to 35 characters long.
 
 
 

164 lines
7.4 KiB

00188266 INFO [SMTP2 ] tx:/1N2v0R
00188293 INFO [SMTP2 ] rx:(6) (2F3040030D0A) /0@
00188294 INFO [SMTP2 ] tx:/1N2c287100R
00188349 INFO [SMTP2 ] rx:(6) (2F3040030D0A) /0@
00188350 INFO [SMTP2 ] pump_enable_temp_compensation 0
00188351 INFO [SMTP2 ] tx:/1u24,0R
00188403 INFO [SMTP2 ] rx:(6) (2F3040030D0A) /0@
00188404 INFO [SMTP2 ] tx:/1K0R
00188449 INFO [SMTP2 ] rx:(6) (2F3040030D0A) /0@
00188450 INFO [PipetteModuleV2] <- call success
00188467 INFO [PipetteModuleV2] m82 bfcall kpipette_aspirate(75000) ->
00188478 INFO [PipetteModuleV2] <- call success
00188479 INFO [PipetteModuleV2] start work -> pipette_aspirate
00188480 INFO [SMTP2 ] pump_aspirate 75000
00188481 INFO [SMTP2 ] tx:/1?3
00188503 INFO [SMTP2 ] rx:(12) (2F3060313439393839030D0A) /0`149989
00188504 INFO [SMTP2 ] pump_move_to_nl 224989
00188505 INFO [SMTP2 ] tx:/1T
00188518 INFO [SMTP2 ] rx:(6) (2F3060030D0A) /0`
00188519 INFO [SMTP2 ] tx:/1CR
00188527 INFO [SMTP2 ] rx:(6) (2F3060030D0A) /0`
00188528 INFO [SMTP2 ] tx:/1V31900A224989R
00188552 INFO [SMTP2 ] rx:(6) (2F3040030D0A) /0@
00188553 INFO [PipetteModuleV2] zm_move_to 300 10
00190998 INFO [SMTP2 ] tx:/1TR
00191013 INFO [SMTP2 ] rx:(6) (2F3060030D0A) /0`
00191014 INFO [SMTP2 ] tx:/1TR
00191029 INFO [SMTP2 ] rx:(6) (2F3060030D0A) /0`
00191030 INFO [PipetteModuleV2] <- end work
00191069 INFO [PipetteModuleV2] m82 bfcall kpipette_zmotor_move_to(0) ->
00191089 INFO [PipetteModuleV2] <- call success
00191090 INFO [PipetteModuleV2] start work -> pipette_zmotor_move_to
00191091 INFO [PipetteModuleV2] zm_move_to 0 1500
00191466 INFO [SMTP2 ] tx:/1TR
00191481 INFO [SMTP2 ] rx:(6) (2F3060030D0A) /0`
00191482 INFO [PipetteModuleV2] <- end work
00191646 INFO [PipetteModuleV2] m82 bfcall kmodule_get_reg(4102) ->
00191647 INFO [PipetteModuleV2] <- call success
00191664 INFO [PipetteModuleV2] m82 bfcall kpipette_zmotor_read_zero_point_state ->
00191665 INFO [PipetteModuleV2] <- call success
00191682 INFO [PipetteModuleV2] m82 bfcall kpipette_zmotor_read_zero_point_state ->
00191683 INFO [PipetteModuleV2] <- call success
00192687 INFO [PipetteModuleV2] m82 bfcall kpipette_zmotor_enable(1) ->
00192689 INFO [PipetteModuleV2] <- call success
00192725 INFO [PipetteModuleV2] m82 bfcall kpipette_zmotor_move_to(50) ->
00192727 INFO [PipetteModuleV2] <- call success
00192728 INFO [PipetteModuleV2] start work -> pipette_zmotor_move_to
00192729 INFO [PipetteModuleV2] zm_move_to 50 1500
00192891 INFO [SMTP2 ] tx:/1TR
00192906 INFO [SMTP2 ] rx:(6) (2F3060030D0A) /0`
00192907 INFO [PipetteModuleV2] <- end work
00192990 INFO [PipetteModuleV2] m82 bfcall kliquid_operation_clear_params ->
00192991 INFO [PipetteModuleV2] <- call success
00193010 INFO [PipetteModuleV2] m82 bfcall kliquid_operation_set_gun_runparams(14,14,0,1000,8000) ->
00193011 INFO [PipetteModuleV2] <- call success
00193027 INFO [PipetteModuleV2] m82 bfcall kliquid_operation_fresh_params ->
00193028 INFO [PipetteModuleV2] liquid_operation_runparam:
00193029 INFO [PipetteModuleV2] pump_acc :14
00193030 INFO [PipetteModuleV2] pump_dec :14
00193031 INFO [PipetteModuleV2] pump_vstart :0
00193032 INFO [PipetteModuleV2] pump_vstop :1000
00193033 INFO [PipetteModuleV2] pump_vmax :8000
00193034 INFO [PipetteModuleV2] zmotor_posmin :0
00193035 INFO [PipetteModuleV2] zmotor_posmax :0
00193036 INFO [PipetteModuleV2] zmotor_vmax :0
00193037 INFO [PipetteModuleV2] lld_enable_record_data :0
00193038 INFO [SMTP2 ] tx:/1L14,14R
00193058 INFO [SMTP2 ] rx:(6) (2F3040030D0A) /0@
00193059 INFO [SMTP2 ] tx:/1N2v0R
00193086 INFO [SMTP2 ] rx:(6) (2F3040030D0A) /0@
00193087 INFO [SMTP2 ] tx:/1N2c319000R
00193191 INFO [SMTP2 ] rx:(0) ()
00193212 INFO [SMTP2 ] tx:/1N2c319000R
00193316 INFO [SMTP2 ] rx:(0) ()
00193337 INFO [SMTP2 ] tx:/1N2c319000R
00193441 INFO [SMTP2 ] rx:(0) ()
00193462 ERRO [PipetteModuleV2] [366] do m_smtp2.pump_set_vstop(liquid_operation_runparam.pump_vstop) fail, error kdevice_is_offline(1105)
00193463 INFO [PipetteModuleV2] <- call fail,kdevice_is_offline(1105)
00240684 INFO [PipetteModuleV2] m82 bfcall kmodule_get_reg(4002) ->
00240685 INFO [SMTP2 ] tx:/1?31
00240710 INFO [SMTP2 ] rx:(6) (2F3040030D0A) /0@
00240711 INFO [PipetteModuleV2] <- call success
00240779 INFO [PipetteModuleV2] m82 bfcall kmodule_get_reg(4102) ->
00240780 INFO [PipetteModuleV2] <- call success
00240797 INFO [PipetteModuleV2] m82 bfcall kpipette_zmotor_read_zero_point_state ->
00240798 INFO [PipetteModuleV2] <- call success
00240813 INFO [PipetteModuleV2] m82 bfcall kpipette_zmotor_move_to_zero_point_quick ->
00240830 INFO [PipetteModuleV2] <- call success
00240831 INFO [PipetteModuleV2] start work -> pipette_zmotor_move_to_zero_point_quick
00240832 INFO [PipetteModuleV2] zm_move_to 0 1500
00241002 INFO [PipetteModuleV2] leave from zero
00241003 INFO [TMC5130 ] moveToEnd 1 100
00241019 INFO [PipetteModuleV2] move to again
00241020 INFO [TMC5130 ] moveToEnd -1 200
00241047 INFO [SMTP2 ] tx:/1TR
00241062 INFO [SMTP2 ] rx:(6) (2F3060030D0A) /0`
00241063 INFO [PipetteModuleV2] <- end work
00241096 INFO [PipetteModuleV2] m82 bfcall kpipette_zmotor_read_zero_point_state ->
00241097 INFO [PipetteModuleV2] <- call success
00242225 INFO [PipetteModuleV2] m82 bfcall kpipette_zmotor_enable(1) ->
00242231 INFO [PipetteModuleV2] <- call success
00242247 INFO [PipetteModuleV2] m82 bfcall kpipette_zmotor_move_to_zero_point_quick ->
00242250 INFO [PipetteModuleV2] <- call success
00242251 INFO [PipetteModuleV2] start work -> pipette_zmotor_move_to_zero_point_quick
00242252 INFO [PipetteModuleV2] zm_move_to 0 1500
00242278 INFO [PipetteModuleV2] move to zero
00242279 INFO [TMC5130 ] moveToEnd -1 200
00242305 INFO [PipetteModuleV2] leave from zero
00242306 INFO [TMC5130 ] moveToEnd 1 100
00242322 INFO [PipetteModuleV2] move to again
00242323 INFO [TMC5130 ] moveToEnd -1 200
00242350 INFO [SMTP2 ] tx:/1TR
00242365 INFO [SMTP2 ] rx:(6) (2F3060030D0A) /0`
00242366 INFO [PipetteModuleV2] <- end work
00242400 INFO [PipetteModuleV2] m82 bfcall kpipette_zmotor_move_to(545) ->
00242415 INFO [PipetteModuleV2] <- call success
00242416 INFO [PipetteModuleV2] start work -> pipette_zmotor_move_to
00242417 INFO [PipetteModuleV2] zm_move_to 545 1500
00243001 INFO [SMTP2 ] tx:/1TR
00243016 INFO [SMTP2 ] rx:(6) (2F3060030D0A) /0`
00243017 INFO [PipetteModuleV2] <- end work
00243146 INFO [PipetteModuleV2] m82 bfcall kpipette_zmotor_enable(1) ->
00243156 INFO [PipetteModuleV2] <- call success
00243190 INFO [PipetteModuleV2] m82 bfcall kpipette_init_device ->
00243194 INFO [PipetteModuleV2] <- call success
00243195 INFO [PipetteModuleV2] start work -> pipette_init_device
00243196 INFO [SMTP2 ] pump_set_io1_mode 0
00243197 INFO [SMTP2 ] tx:/1u9,0R
00243215 INFO [SMTP2 ] rx:(6) (2F3040030D0A) /0@
00243216 INFO [SMTP2 ] pump_set_io2_mode 0
00243217 INFO [SMTP2 ] tx:/1u10,0R
00243247 INFO [SMTP2 ] rx:(6) (2F3040030D0A) /0@
00243248 INFO [SMTP2 ] pump_set_tip_type 1
00243249 INFO [SMTP2 ] tx:/1u18,1R
00243279 INFO [SMTP2 ] rx:(6) (2F3040030D0A) /0@
00243280 INFO [SMTP2 ] tx:/1T
00243308 INFO [SMTP2 ] rx:(6) (2F3060030D0A) /0`
00243309 INFO [SMTP2 ] tx:/1CR
00243317 INFO [SMTP2 ] rx:(6) (2F3060030D0A) /0`
00243318 INFO [SMTP2 ] tx:/1N2V2552000ZR
00243341 INFO [SMTP2 ] rx:(6) (2F3040030D0A) /0@